Wrap Care
See Wrap Use Guide for instructions on how to clean the exterior of the Wraps.
The
interior
of the Wrap can be cleaned once a month by emptying 2-3 bottles of
16-24 oz. hydrogen peroxide (2%) into the ice box of the Control Unit and running the
system for five minutes. Fill the ice box approximately four inches deep with hydrogen
peroxide. Leave the ice box lid open. Adjust the pressure setting to the “no pressure”
mode. Adjust the temperature to its coldest setting (3 snowflakes). Set the timer for
five minutes and press the start button.
Caution:
Leave the ice box lid open when using hydrogen peroxide to clean the interior
of the Wrap. If the ice box is sealed shut, hydrogen peroxide can form a gas that could
build excess pressure inside the unit and potentially damage the system.
Wrap Storage
Do not store Wraps in extreme cold temperatures (below 33F). To store, coil the
Connector Hose, and fold the Wrap neatly on its natural folds. Handle your
Wraps with care.
Caution:
Do not store Wraps inside a plastic bag or other airtight container as it may
promote mold growth.
Bag Care and Storage
Wipe clean with mild soap and water, air dry.
Caution:
Store the Game Ready System out of the reach of children and pets.