Machine Shut Down
In case of a battery failure simultaneous with a power failure the
machine will shut down and a segment on the Time Display will
be
fl
ashing. The settings and accumulated values are stored by the
machine. When the power returns the machine will perform a recovery,
see Recovery from Machine Shut Down on page 3:29.
Return the Blood During Machine Shut Down
During a machine shut down, it is possible to manually return the
blood to the patient.
1. Clamp the arterial blood line and disconnect it from the patient.
2. Connect the arterial blood line to the rinse-back solution and
remove the clamp.
3. Open the cover of the blood pump and turn the blood pump
manually.
4. When the required amount of blood has been returned to the
patient, clamp the venous blood line and disconnect it from the
patient.
WARNING
During manual procedure to return blood to patient in power failure,
the operator assumes responsibility for visually monitoring all safety
parameters which the machine cannot monitor in a power failure
(e.g. air detection).
WARNING
3:28
AK 96
®
Operator's Manual - Operating the Machine
HCEN12239 Revision 12.2010
Program version 3.xx
Содержание AK 96
Страница 2: ......
Страница 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 281: ...This page is intentionally left blank 11 16 AK 96 Operator s Manual BPM HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...