Fluid Part Component Details
1.
Safety Couplings for the Dialysis Fluid Tubes
The dialysis
fl
uid tubes must be connected to the safety
couplings during the initial part of the function check of the
machine and when disinfection/rinse programs are running.
To correctly attach
the dialysis
fl
uid tube to the safety
coupling;
press and hold the button on the dialysis
fl
uid
tube connector
while attaching it to the safety coupling.
Release the button and simultaneously
push the connector
into place until it clicks in
. The small button positioned
just below the safety coupling will now be depressed
which means that the connector is correctly attached to the
machine.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Machine Description
2:19