→ → → CAUTION
Preventive inspection, maintenance and calibration of the
AK 96 dialysis machine shall be performed by a fully trained
authorized service technician according to the Maintenance Manual
in the AK 96 Service Manual which can be ordered from your
Gambro representative. It is mandatory for preventive maintenance
to be performed at least every other year. Yearly maintenance
is recommended. The interval between preventive maintenance
procedures might differ due to operating environment variations.
The AK 96 dialysis machine is in compliance with certain
requirements concerning patient leakage current from the dialysis
fluid in accordance with international standards and regulations.
When a central venous catheter is used, and the tip of the
catheter is close to the heart, it is however necessary to take
extra precautions to minimize the risk of arrythmia due to leakage
currents. For these treatments it is necessary to connect the
potential equalization conductor between the AK 96 dialysis
machine and the potential equalization bus bar in the electrical
installation. Potential equalisation also has to be used when legal
requirements of the installation place requires it.
To minimize the leakage currents from other electrical equipment it
is recommended to place such equipment outside the patient area.
Any equipment within the patient area shall fulfil the IEC 60601-1
and IEC 60601-1-1 standards and be a part of the potential
equalization. One way to minimize the leakage currents from
equipment within the patient area is to electrically isolate it. Make
sure leakage current values are below respective limit required by
CF type applied parts. Check with authorized technician.
If your clinic/hospital uses a central venous catheter during
treatments please make sure your machine is equipped with a
potential equalization connection. If not please contact your local
Gambro Service Technician for further assistance.
The AK 96 dialysis machine is not suitable for use in the presence
of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide.
CAUTION
1:12
AK 96
®
Operator's Manual - Before you get started
HCEN12239 Revision 12.2010
Program version 3.xx
Содержание AK 96
Страница 2: ......
Страница 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 281: ...This page is intentionally left blank 11 16 AK 96 Operator s Manual BPM HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Страница 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...