If this cable is replaced, remember that the yellow-green earth wire must be at least 40 mm longer than
the live and neutral wires. This will ensure that if the cable is jerked, the last wire to disconnect will be the
earth wire, thus maintaining electrical protection at all times.
This operation and any other technical intervention will be carried out by the corresponding Technical
Assistance Service.
The bath will be always connected to a fixed cable trough.
3. Protection
The bath is provided with equipotential connection between its metallic elements. This connection,
located on the motor base plate, must be connected to the bathroom’s equipotential installation. The
2
2
connecting wire may have a section from anywhere between 2,5 mm and 6 mm .
The electrical installation of the bath must with a protected by differential switch, whose current of
maximum deviation will be the indicated in the country where the bath is mounted as well as with a
magneto-thermal circuit breaker appropriate for the consumption of the bath (see consumption on the
technical specifications). The latter will be provided with an omnipolar disconnection and with an opening
distance of at least 3 mm. Both circuit breakers must be located outside the protection volumes of the
bathroom.
5. Wiring diagrams
6. Technical specifications:
• Supply voltage: . . . . . . . . . . . . . .220... 230V~; 50Hz
WATER PUMP
650W
2,9A
IP55
Wattage rating
Rated current
Protection
Installation manual