chapter 5 - page 4
gako unguator Jar
Error
Troubleshooting
1
The jar connected to the gako unguator EMP is
not a gako unguator jar.
The jar has to be substituted for a gako unguator
jar.
2
The gako unguator Jar Lid was screwed alsant
onto the jar and the stirrer cants on automatic
insertion.
Open the jar and screw the lid on correctly.
3
At
fi
rst time extraction the ointment leaks out in
a gush.
Diminish air before the
fi
rst time extraction.
4
Liquid leaks out on the bottom of the jar.
• The bottom of the jar was not completely
covered with foundation all the way to the rim
of the jar bottom. Ensure to cover the sealing
lip of the bottom with foundation.
• After using the jar in a water bath, water
accumulated in the cavity on the bottom side
of the jar. The water cannot penetrate into the
jar itself through the sealing lip, remove the
excess water with a paper towel.
5
The jar does not attach into the lift arm.
• The jar is not an original gako unguator Jar.
Always use an original gako unguator Jar.
• Please observe the correct handling of the
quick-lock coupling (see description below)
Hold the gako logo, on top of the jar, to the front. Load the stirring shaft directly from the front
into the lift arm. Push the mixing unit up and with a slight turn counterclockwise (about a
quarter turn)
fi
x the mixing unit into the lift arm.
tab.: Operation Errors / Jar
pic.: quick-lock Coupling