Responsibility of the installer.
It is the responsibility of the installer to inspect the product prior to installation to ensure the unit is free of defect or
damage. In the event of a problem the unit must not be installed.
This product complies with internationally recognised safety standards. However you are responsible for any local
codes that may apply.
This product is designed to be installed by a licensed tradesperson. Licensed plumbers and electricians should be
employed to ensure a proper installation. Installers assume all liability for the correct installation procedures.
Only accessories authorized by the manufacturer should be used with this product.
Risk of electric shock.
Connect only to a separate circuit protected by a 30mA RCCD.
The unit must be connected to a RCCD. This should be tested on a routine basis. To test the RCCD push the test
button, the RCCD should interrupt power. Push the reset button, power should be restored.
If the RCCD fails to operate in this manner there is a ground current flowing indicating the possibility of electric
shock. Do not use this unit. The unit should be corrected by a qualified representative before any further installation
is undertaken.
Risk of electric shock
. Spa motors where fitted should be connected to a separate 30mA RCCD protected outlet.
Note: The unit must not be fitted in the bathroom area unless protected by an RCCD.
Classification & technical data.
Fixed class 1 installation
Type B applied part
Not suitable for use in the presence of flammable anaesthetic mixture with air or oxygen or nitrous oxide.
Rated supply 230v, 240W 60Hz refer to label.
Maximum patient weight: 22 stones . 150kg (140 Kg transfer systems)
Maximum weight of bathtub, patient and water 420Kg
Maximum weight of complete transfer system . 33kg
Suitable for intermittent use only - 2 minutes on - 18 minutes off.
The power pack is low voltage and runs at 24v DC . Immediately after installation and prior to use it is vital that the
bath is connected to a mains supply for at least 24 hours before initial use to ensure the batteries are fully charged.
Failure to comply with this requirement could result in the user not being able in the event of a mains power failure.