To get the most out of your Malvern bath we recommend that the unit is cleaned and maintained on a daily basis.
The chrome seat tube should be cleaned using a silicone based cleaning product (such as Mr Sheen)
Spray a small amount on the tube when the seat is fully raised and wipe with a dry soft cloth. This will avoid any
build up or soap, water or atmospherics.
The bath can be cleaned with a medical disinfectant, especially if the product is to be used with multiple users.
Do not use abrasive type cleaners. Oven cleaners, scouring powders or bleach on this product.
Code 1: Lift actuator fault
:
One long flash followed by 1 short flash
----------
-----
---------- -----
Code 2:
Traverse actuator fault
:
One long flash followed by 2 short flashes
----------
----- -----
---------- ----- ----
Code 3:
Leg Lift actuator fault
:
One long flash followed by 3 short flashes
----------
----- -----
---------- ----- ----
Code 4: Bath lift actuators fault
:
One long flash followed by 4 short
flashes
----------
----- ----- ----- -----
---------- ----- ----- ----- -----
If any of the above fault codes are indicated on the status display please contact our service department.
1.
The seat will not traverse — make sure the bath is fully lowered.
2.
The seat will not raise — Check that the system has mains power (green light on status display)
3.
Bath will lower but not raise — Check that the system has mains power (green light on status display)
4.
Flashing blue light on status display — disconnect mains power, wait 5 minutes, re apply power.
For any other issues please contact the service department.