background image

 

E006

                        

 

 

Page 34                                                                                                                                                             

8 T

IPS AND 

A

SSISTANCE

  

 

 

Table 4 Frequent Asked Questions (FAQs) 

Questions 

 Possible Cause/ Solution 

“STP” LED 

does not Flash. 

 

Device might be in Sleep Mode or “RUN” Mode.

 

 

Device date and time are not set by user. 

How to change the date format?  

 

On power up condition, user can select the date format.  After 
selection 

it cannot be altered in data logger. 

 

The default format is dd-mm-yy 

When to set the RTC in Device?

 

 

 

 

 

After device activation, at power up condition it is required to 
set the RTC. 

 

User can also adjust the RTC from main menu options. 

Device is not connected in 
software application. 

 

USB Symbol must be shown on Display. During insert of USB 
cable, 

“STP” and “RUN” 

LEDs Flash together. 

 

Try to reconnect Micro USB data cable. 

 

Micro USB cable might be faulty. Replace the cable. 

Display Shows “

SnC

” message.

 

 

Sensor cable might not be inserted properly. 

Display Shows “Err” message.

 

 

Sensor  cable  might  be  broken  or  temperature  is  out  of 
measuring range. 

What happens if Sensor is not 
connected after device wake up? 

 

Recording  of  temperature  will  be  delayed  by  the  time  for 
which sensor is not connected. 

E.g.,  Once  device  wake  up  and  RTC  set,  delay  counter  starts.  If 
sensor is removed from data logger, 

within 5 min “SnC”

 message 

is  displayed and delay counter  shows  00:05(Hr: Mn). If sensor is 
connected after 10 minutes, then delay counter resumes from the 
last count and the recording is delayed by 10 minutes. 

For how much time, the display 
remains ON after device 
activation? 

 

Once  the  device  is  activated,  the  display  remains  ON  for  10 
minutes, then it will off. 

 

The  display  is  normally  off  to  save  the  battery  life  when  no 
activity on device. 

How to set the time of device if it 
has offset from local time? 

 

User can adjust the time using “

t

ME” 

menu (

section 6.4.9

). 

How to set the date of device if it 
has offset from local date? 

 

User can adjust the time 

using “

dtE

” 

menu (

section 6.4.8

). 

What are the conditions for 
buzzer activation?  

 

Please refer the Buzzer operation conditions in  

section 6.7 

How to deactivate the Buzzer for 
some time in continuous alarm 
condition? 

 

Buzzer can be deactivated by pressing 

Function

 and 

Enter

 

key at same time for alarm acknowledgement delay. 

How to activate pause function? 

 

Press  “Function”  and  “Up”  key  for  5  sec  simultaneously  to 

activate the pause function. (Refer 

section 6.8

 for details) 

What to do if user wants to view 
updated temperature quickly on 
display, after changing the 
temperature? 

 

User can press “Up” and “Enter” key simultaneously for 3 to 4 

times to get quick update view of temperature reading. 

 

Содержание LM -XS PRO E006

Страница 1: ...rature Data logger Model No 99963 English 2021 Rev 1 1 G Tek Corporation Pvt Ltd 3 mahavir estate karelibaug vadodara 390 018 tel 91 265 2461912 email info gtek india com url www gtek india com OPERAT...

Страница 2: ...unting of LM XS Pro E006 Data Logger 10 3 4 Sensor Connection of LM XS Pro E006 Data Logger 10 4 LIST OF ABBREVIATIONS 11 5 PRODUCT DESCRIPTION 12 5 1 Status LEDs 12 5 2 Display LCD 12 5 3 Key functio...

Страница 3: ...26 6 8 Paused Function 27 6 9 Reading out Data 27 6 9 1 Connecting with the Software Application 27 6 9 2 Generating PDF report 28 6 9 3 PDF Report Explanation 30 6 9 4 Definition of Important Terms...

Страница 4: ...nu Sequence 17 Figure 11 Alarm Set Point High 18 Figure 12 Alarm Set Point Low 18 Figure 13 Alarm Acknowledge Delay 19 Figure 14 History Menu to view Min Max Temperature 20 Figure 15 View History data...

Страница 5: ...and personnel also Do not use the product if there are signs of damage to the housing There are no user serviceable parts inside For any defect please consult the factory or the dealer from where you...

Страница 6: ...PC To download data to generate a report in pdf format and to export data in csv format for further use 2 2 Technical Data Table 1 Technical Specifications Model LM XS PRO E006 General Remote Sensor T...

Страница 7: ...larm calendar clock duration delay counter Alarm high and Low Alarm marker Bell symbol REC Pause indication and Current reading with measurement unit Memory Size 30 days overview on the display PDF re...

Страница 8: ...ing IEC60529 IP 64 External Sensor not Plugged in Impact Resistance 5 drops from 1 meter onto concrete floor at room temperature with battery in place Device does not get damaged and there is no loss...

Страница 9: ...d be opened immediately and the device be examined If the device is found damaged it should not be operated and the local representative contacted for instructions Ensure that all accessories and docu...

Страница 10: ...chanical Dimensions of LM XS Pro E006 Data Logger Figure 2 Overall dimensions of LM XS Pro E006 Data Logger Overall Dimensions Dimension L x W x H mm 128 x 60 x22 approx Mounting Screw Mounted Weight...

Страница 11: ...crew mounting 3 4 Sensor Connection of LM XS Pro E006 Data Logger Figure 4 External Sensor attachment Figure 5 External Sensor jack Inserted Sensor will be inserted in the LM XS Pro E006 data logger u...

Страница 12: ...Time setting dd Date MM Month YY Year Hr Hour Mn Minute ASH Alarm Set Point High ASL Alarm Set Point Low ACd Alarm Acknowledge Delay ALA Alarm Acknowledged HSt History YES Yes ALH Alarm History dIF D...

Страница 13: ...position and description of each segment is shown in figure 6 Figure 6 LCD Display format 1 OK NOK X symbol a If any time in last 30 days alarm limits are crossed the symbol X NOK will be turned on an...

Страница 14: ...tart indicator when first time logger is started by setting calendar it will wait for 10 minutes to start logging the data During these 10 minutes time only this sand clock symbol will be on This symb...

Страница 15: ...h will start after 10 minutes of device activation During this delay time sand clock symbol will be visible and STP LED will be flashing Once the recording of data is started STP LED turns off and RUN...

Страница 16: ...ailable calendar formats dd mm yy or mm dd yy by following the sequence as shown in figure 7 Once selected date format cannot be changed throughout the life of data logger If the user has not set the...

Страница 17: ...21 in year it will be autocorrected as 28 02 21 dd mm yy 6 2 View Min Max and Current Data As described in key functions section 5 3 by pressing Function and Up key at the same time for 1 sec in norm...

Страница 18: ...f LM XS Pro E006 data logger using main menu In this menu the user can view Alarm set point high low along with its duration Alarm acknowledgement delay history data alarm history data and device info...

Страница 19: ...11 Alarm Set Point High 6 4 2 ASL Alarm Set Point Low In this menu Alarm setpoint Low along with its alarm delay can be seen which is preset at 0 5 C and 1 hour User can only view this parameter Alarm...

Страница 20: ...nge for HR and MN is 00 to 23 and 00 to 59 respectively 6 4 4 HST History Menu The user can see the history data of min max values for last 30 days using history menu In this menu the user can choose...

Страница 21: ...e display will be showing the date of 01 day alarm trigger time In case of alarm history data for Min Max values along with its duration and sensor open broken condition duration if any in sequence up...

Страница 22: ...ew terminates automatically if there is no alarm data to show in last 30 days Figure 16 View Alarm History data in last 30 days Note Normal history and Alarm history viewing can be terminated manually...

Страница 23: ...rature unit CEL C FAH F for viewing by following the steps shown in figure 18 User can view the data in degree Fahrenheit but data is stored in degree Celsius only Figure 18 Select Unit for Temperatur...

Страница 24: ...hanged will be displayed Every time date is changed a date change Tag is logged and it can be seen by downloading the data using LmView XS E006 software application Figure 19 Date setting menu in dd m...

Страница 25: ...after 10 minutes of device RTC set on power up The Configuration parameters are prefixed as per WHO Specification reference E006 TR06 3 Issue date September 8 2010 and user cannot change them from de...

Страница 26: ...high low condition has occurred R Alarm Reset Alarm high low condition is restored T Time set old time Hr Mn User has updated the current time P Pause Data User has paused data statistics of alarm min...

Страница 27: ...on Buzzer will be activated in following conditions 1 Alarm High Low In case temperature goes beyond alarm set point high low after alarm delay high low buzzer will be activated for 1 sec at every 1 h...

Страница 28: ...reading after paused function is activated by pressing Function and UP key for 5 seconds This will be indicated by PAUSE message along with pause time counter and sand clock on the display Prefixed Pa...

Страница 29: ...ss readings The pop up window will appear as shown in figure 22 Select the appropriate file path and batch file name for saving the data summary and press button Figure 22 Selecting file location for...

Страница 30: ...After that and generate the PDF report of data summary by clicking on icon A pop up window will appear for selecting the file path and filename of report to be generated as shown in figure 24 Figure 2...

Страница 31: ...and last record time 5 Statistical Summary Statistical analysis of the total logged data 6 Data Summary Shows max 120 days summary in table Each row consists of a day summary Date Entry of date is in...

Страница 32: ...high trigger time and its cumulative time Sensor connection error trigger time and cumulative time Figure 27 Sample PDF report part 3 The gray band in graph shows area inside alarm low and high limit...

Страница 33: ...hat the data points tend to be very close to the mean a high std deviation indicates that the data points are spread out over a large range of values 3 Cumulative time Actual cumulative daily time dur...

Страница 34: ...is indicated by a low battery symbol the data logger should be replaced within 15 days when this symbol appears Dispose or recycle the battery in accordance with your local regulations Do not expose t...

Страница 35: ...ensor is removed from data logger within 5 min SnC message is displayed and delay counter shows 00 05 Hr Mn If sensor is connected after 10 minutes then delay counter resumes from the last count and t...

Отзывы: