E006
Page 27
3.
Sensor Broken:
When sensor cable is broken or the reading goes out of device
temperature range, device display will show
“Err”
message on the device display. Buzzer
operation will be same as described in Sensor Open condition
, except showing “Err”
message on display.
6.8 Paused Function
➢
The Pause function is useful, for example, when you wish to temporarily remove the sensor
from the monitored location to inspect goods, but you do not wish to trigger an alarm due to
your handling.
➢
This allows the user to review the current statistics or clear an alarm without causing a false
alarm or statistic while handling the data logger.
➢
The data logger is configured to halt processing of temperature data for alarms and min/max
statistics for a period of two/ three temperature reading after paused function is activated
by pressing
“
Function
”
and
“
UP
”
key
for 5 seconds. This will be indicated by
“PAUSE”
message along with pause time counter and sand clock on the display.
➢
Prefixed Pause time for the data logger is 15 minutes. After pause time out, the data logger
will resume to normal operation,
“PAUSE”
message will disappear and
“REC”
will be
displayed.
➢
If sensor is open/ broken during the pause delay,
“SNC”
or
“Err”
message will be seen on the
display along with pause time counter and message.
➢
If the data logger is in alarm condition and user has paused the logging, then the pause time
is excluded from the alarm trigger time and duration calculation.
6.9 Reading out Data
6.9.1 Connecting with the Software Application
Displaying a measurement data report
➢
Connect the LM-XS Pro E006 data logger to Windows PC via the USB port, as shown in
figure 21.
➢
After connecting the data logger with PC, display remains ON and it shows USB port symbol
along with the other values.