E006
Page 25
➢
Tag event is logged again when alarm high/low condition is restored in temperature
measurement.
➢
When Date is set a special tag event with time stamp is logged in data records.
➢
Time setting Tag event is logged when the user adjusts the time in device running condition.
➢
If sensor is broken or not inserted properly, tag event is logged as sensor open.
➢
When the user has paused the data logging, pause event is logged and after 15 minutes Start
after pause event is logged.
Tag ID Tag Event
Description
D
Date Set
User has set the date.
E
Sensor Open
Sensor open or broken (Reading goes
beyond specified temperature range for
the sensor)
A
Alarm Triggered
Alarm high/low condition has occurred.
R
Alarm Reset
Alarm high/low condition is restored.
T
Time set: old time Hr: Mn User has updated the current time.
P
Pause Data
User has paused data statistics of alarm,
min/max for 15 minutes.
S
Resume From Pause
Normal recording is resumed after 15
minutes.
Note: Number of data records get reduced according to number of Tag data events.
Tag Event(s)
can be seen in software application and generated pdf report after downloading data.
6.6 Displaying of Reading in Normal and Alarm Condition
There are four possible circumstances of displaying temperature reading for data logger:
1)
Reading is within the set point high/ low.
-
OK sign, battery, reading and unit will be seen on display.
2)
Reading is outside the setpoint high/ low for time less than alarm high/low delay.
-
OK sign, battery, reading with UP/ Down arrow and unit will be seen on display.