Page 4
2
S
PECIFICATIONS
2.1. Use
LM
Pro
T051 User Programmable Temperature Data logger
is an Internal sensor data logger
meeting the requirements of
WHO PQS E006/TR05.1.
It can store up to 50000 data records.
User can fully configure parameters like
–
Start delay, alarm high/low with corresponding
alarm delay, alarm event type- single/cumulative, temperature unit, store interval etc. User
can also set the information like, username and job description. Software automatically sets
the UTC time zone at the time of configuration of device. User can also download data from
multiple data loggers in single file. User can generate report in pdf format or export data as a
csv file. Data loggers come with individual NABL traceable calibration certificate.
The temperature readings are monitored and saved throughout the entire duration of
measurement program.
LMViewE051 software needs to be installed in PC To download data to generate a report in
pdf format and to export data in csv format for further use.
2.2 Technical Data
Table 1 Technical Specifications
Model
LM
Pro
T051
General
Integrated Sensor
Thermistor - 10K NTC
Temperature
Measurement Range
-30 °C to + 60 °C (-22 °F to +140 °F)
Accuracy
± 0.5 °C for the range -20 °C to + 30 °C
± 0.7 °C otherwise
Resolution
0.01 °C - Display and
Storage
Unit of Measurement
Data in °C. User has an option to view the data in °F
Calibration
Each device accompanies NABL (ISO/IEC 17025) traceable certificate
Alarm
Visual
Alarm Settings
Fully user programmable four alarms with alarm type(High/Low), event
type(single/cumulative) , alarm delay(Hr:Mn) selection.
Response Time
T
90
< 20 minutes as per EN12830:1999
Logging Interval
Logging interval is user programmable from 5 seconds to 18 hours
Delayed Start Option
Yes. User can program start delay from 0 second to the required time by
selecting date and time.
Power Requirement
Battery
User Replaceable 3.0 V 225 mAh; CR2032 Panasonic (or Equivalent) Coin
Cell Battery;