9
D1020
- SIL 2 Powered Isolating Driver Smart-Hart compatible
G.M. International ISM0012-14
Configuration
Screwdriver
for Trimmers
Ch1 and Ch2
1 x 0.5 mm
Calibration Trimmer Ch1
Span Trimmer
CH1
Calibration Trimmer Ch2
Side B Panel View
Turn the trimmer clockwise to
Increase Output value or turn
the trimmer counterclockwise
if you want to decrease Output
value.
Side A Panel View
Span Trimmer
CH2
Calibration instruction and Troubleshooting
In case a calibration check or readjustment is necessary the following equipment is required:
Set power supply (PS) at 24.0 V at terminals 3 (+) and 4 (-), the calibrator (CAL) in the function “mA” 20 mA range at terminals 1 (+) and 2 (-), the multimeter (V1) for the reading in
DC supply 30 V range at terminals 1 (+) and 2 (-), the multimeter (DMM) for the reading in DC supply 20 mA range at terminals 14 (+) and 15 (-).
Execute the next test to verify D1020S unit channel or execute it twice to verified both D1020D unit channels.
Verify that green led on the front panel is ON.
Connect the unit as shown below:
If the power ON Led is off, check supply voltage, polarity and wiring.
Set calibrator (CAL) at 4.000 mA and verify that the output current, read by multimeter (DMM), is correspondent with a maximum deviation presented in the section “Technical Data”.
Set calibrator (CAL) at 20.000 mA and verify that the output current, read by multimeter (DMM), is correspondent with a maximum deviation presented in the section “Technical Data”.
Set calibrator (CAL) at 20.000 mA and power supply (PS) at 20.0 V; verify that multimeter (V1) reads input voltage as presented in the section “Technical Data”.
If the maximum deviation is exceeded, repeat the board calibration and proceed in the following way:
Set the calibrator (CAL) at 20.000 mA and regulate the current, read by multimeter (DMM), by turning the span trimmer of the unit (see “Configuration” section) until reading 20.000
mA ± 5 µA value.
To execute a linearity test: set the calibrator (CAL) at 4.000, 8.000, 12.000, 16.000, 20.000 mA and verify current output is correspondent with
a maximum deviation presented in the section “Technical Data”.
Adjustable Power Supply Range 0-35 V Resolution 0.1 V,
Accuracy 1% Output Capability 200 mA
Current Meter range 20 mA, Resolution 1 µA Accuracy 0.01%
or Voltage Meter Range 10 V, Resolution 1 mV Accuracy 0.01%
Voltage Meter range 30 V, Resolution 1 mV Accuracy 0.05 %
Current Calibrator range 20 mA, Resolution 1 µA Accuracy 0.01 %
Agilent
E3611A
Agilent
34410A
Fluke
175
Yokogawa
7651
Equipment Required
Instrument
PS
DMM
V1
CAL
Identification
PS
Supply
=
_
+
V1
CAL
DMM
mA
mA
MODEL D1020D
=
=
=
14
15
3 +
4 -
1
2
=
=
=
10
11
5
6
=
=
=
=
In 1
V1
CAL mA
Out 1
+
DMM
mA
Out 2
7 +
8 -
In 2