Chapter 5
Maintenance
Confidential
Copying Prohibited
5-53
5.9.5.2
Checking the enclosure leakage current (current that flows from the enclosure to the
earth terminal of the power outlet)
(1) Measurement wiring diagram
(2) Enclosure leakage current measurement points
Measurement point A
Optional location on enclosure; or, if enclosure is insulated,
metal foil attached to enclosure
Measurement point B
Earth terminal of power outlet
Place the MD in between measurement point A and B.
(3) Enclosure leakage current measurement conditions
It is necessary to measure the enclosure leakage current under two conditions: normal
condition and single fault condition.
(4) How to reproduce a single fault condition
You can reproduce a single fault condition by connecting the conversion plug to the power
outlet with the earth wire not connected.
How to change the phase
You can switch the phase by inserting the conversion plug into the power outlet upside
down.
(5) Acceptance standards
Check that the maximum current value meets the following judgment criteria.
Normal condition
0.1 mA or less
Single fault condition
0.5 mA or less
VS-1500
Содержание VaSera VS-1500
Страница 12: ...Confidential Copying Prohibited x ...
Страница 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Страница 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Страница 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Страница 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Страница 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Страница 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Страница 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...
Страница 213: ...Chapter 12 Information of the electromagnetic compatibility EMC 12 6 Confidential Copying Prohibited ...
Страница 216: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 4R0100090 200906 ...