5. Cleaning, Sterilization and Disinfection
Clean, sterilize and disinfect the HLX-561 and accessories as follows.
Cleaning
Use a clean, soft cloth dampened with alcohol or a weak
acid/alkalescent/neutral cleanser. Clean the HLX-561 and accessories
with the dampened cloth, taking care that the cleaning solution does not
enter into any connectors or get inside the HLX-561.
Do not use cleaning agents containing organic solutions, thinner,
toluene and benzene. Such chemical solutions may damage the resin
casing.
Sterilization and Disinfection
For Disinfection, wipe the HLX-561 and accessories using a gauze or
sanitary cotton moistened with inert soap or alcohol. Do not moisten
allow moisture inside of the connector to avoid contact failure.
For sterilization, use a gas sterilization method. Sterilization methods
utilizing high temperatures such as an autoclave will damage the
HLX-561 and accessories.
If sterilizing the room utilizing spray chemical solution methods, take
proper measures to ensure the chemical solution will not contact the
inside of the HLX-561 and connectors.