4
Cleaning and disinfecting
Cleaning and disinfecting
The iViz ultrasound system consists of the system (tablet), accessory case, and transducer.
This section includes instructions for cleaning and disinfecting the iViz ultrasound system. The system must
be cleaned and disinfected after each exam. It is important to follow this cleaning and disinfection process
without shortening or skipping steps. Only use cleaners and disinfectants approved by FUJIFILM SonoSite on
the iViz system. Be sure to observe the solution strength and duration requirements detailed in the cleaning
and disinfection procedures.
Included in this section are instructions for two levels of cleaning and disinfection for both the transducer and
the system. The level of cleaning and disinfection required depends on the type of tissue the system comes
into contact with during an exam. For more information, see
“Cleaning and disinfecting the system and
The materials used in the iViz system are designed and tested to work with the cleaners and disinfectants listed
in this chapter. The iViz system will not be damaged if cleaned according to the approved instructions and with
the approved cleaners and disinfectants. Before using a disinfectant, confirm that it is appropriate for your
facility’s use. Verify expiration dates, concentration, and efficacy of chemicals (for example, a chemical strip
test).
WARNINGS
Ensure that cleaning and disinfecting solutions and wipes are not expired.
Some cleaners and disinfectants can cause an allergic reaction to some individuals.
Transducer
System (tablet)
Accessory case
Содержание SonoSite Edge
Страница 1: ...Cleaning and Disinfecting FUJIFILM SonoSite Products User Guide...
Страница 24: ...18 Compatible high level disinfectants compatible transducers...
Страница 46: ...40 Compatible high level disinfectants transducers...
Страница 68: ...62 Compatible high level disinfectants transducers...
Страница 88: ...82 Compatible high level disinfectants transducers...
Страница 110: ...104 Compatible high level disinfectants transducers...
Страница 150: ...144 Compatible high level disinfectants transducers...
Страница 151: ......
Страница 152: ...P22043 01 P22043 01...