5.1 Installing and Connecting the Equipment
Prepare and inspect the system according to the VP-7000 Operation Manual, “Chapter 6
Preparation and Inspection of the System”.
WARNING
• If any peripherals not described in the VP-7000 Operation Manual, “2.2 Equipment Using
in Combination” are used, this product may not function properly and it may cause damage
to the peripherals or injury to patients or physicians.
• This product conforms to the EMC standard (EN 60601-1-2: 2007). However, the radio waves
radiated form this product may cause medical devices such as a pacemaker to malfunction.
When this product is used for a patient with an active implantable medical device, consult a
cardiovascular specialist and the manufacturer of the active implantable medical device.
Note
• If the battery in the processor discharges, it may impede the clock function displayed
on the monitor (the clock may lose time). If the processor has been left unused for
more than three months, turn ON the power to the processor to charge the battery for
about three hours. (It is not necessary to connect the Endoscope or turn ON the lamp.)
• If the clock displayed on the monitor is not correct, set the date and time again.
5.1.1 Installing the Endoscope and Water Tank
CAUTION
• Before connecting the endoscope, ensure that no foreign matter is caught in the scope
connector section. If any foreign matter is caught in the power supply section, it may cause
malfunction or failure of the devices and may also generate heat. If any foreign matter is
caught, do not touch it directly with your hand but remove it after disconnecting the endoscope.
• Ensure that there is no foreign matter or dirt on the receiving window or on the communication
window before connecting the endoscope. If there is any foreign matter or dirt on the receiving
window or on the communication window, it may cause malfunction or failure of the devices.
• Make sure that the scope connector including the power-receiving section is thoroughly dry
before connecting it to the light source. In addition, make sure that no moisture or foreign
matter (such as metallic fragments, chemical residues, water deposits, grease, dust, and
gauze fibers) adheres to the power-receiving section. If the connector mount is wet or soiled
with foreign matter, it may cause malfunction or failure of the devices.
• Keep metals except for the endoscope away from the power supply section. Otherwise,
metals may generate heat.
• When connecting the endoscope, insert it fully so as to produce no clearance. Do not look
into the connecting part of the endoscope. The light emitted from the light source may cause
damage to the eyes.
BL7000_E-50_897N120551D.indb 1
2016/09/05 14:06:13
Содержание BL-7000
Страница 2: ...ii 897N120551D BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 05...
Страница 10: ...897N120551D 4 BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 06...
Страница 18: ...897N120551D 1 8 Chapter 1 Precautions BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 06...
Страница 32: ...897N120551D 3 8 Chapter 3 Name and Function of Each Part BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 12...
Страница 52: ...897N120551D 6 10 Chapter 6 Method of Use BL7000_E 50_897N120551D indb 10 2016 09 05 14 06 17...
Страница 66: ...897N120551D 8 6 Chapter 8 Troubleshooting BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 19...