897N120551D
1-5
Chapter 1
Precautions
1.2.7
Maintenance
CAUTION
• The equipment will wear out and deteriorate after repeated use over a long period. The
equipment is required to be inspected by specialists once every six months. Also, if any
abnormality is found during clinical use, carefully withdraw the endoscope from the patient
as per instructions provided in the operation manual of the endoscope.
• Do not disassemble or modify the equipment. For details on the inspection, consult your
local FUJIFILM dealer.
1.2.8
Preparation and inspection before use
CAUTION
• Prior to using this product, prepare a spare one to avoid unexpected accidents such as
equipment failure. If a replacement is not available, you may not be able to continue
endoscopic procedures.
•
Make sure to check the equipment before use according to the procedures provided in this
manual, to avoid unexpected accidents, and take full advantage of the equipment’s capabilities.
• In particular, errors in images may cause false diagnosis. If the inspection result shows
any abnormality, do not use the equipment.
1.2.9 Installation of Equipment
CAUTION
•
Since there are ventilation holes on the bottom surface of the light source BL-7000, take
care not to cover the bottom surface with foreign substances.
• Install equipment so that the power cord or connected endoscope cannot be caught by
anything. If the power cord or connected endoscope is caught by something, it may cause
overturning or falling of equipment, disappearance of the endoscopic image or damage to
the patient and/or operator.
1.2.10
Handling
CAUTION
• Keep metals except for the endoscope away from the power supply section. Otherwise,
metals may generate heat.
• Do not disassemble or modify this product. Otherwise, the radiant intensity may exceed
Class I specifications.
•
Clean the equipment in the proper way as specified. Otherwise, the equipment may be
damaged.
• Do not wash the light source with running water or immerse it in disinfectant. The
equipment may be damaged.
• Do not disinfect or sterilize the light source. The equipment may be damaged.
• Wear rubber gloves when handling an endoscope to prevent infection and electrostatic
discharge.
BL7000_E-50_897N120551D.indb 5
2016/09/05 14:06:06
Содержание BL-7000
Страница 2: ...ii 897N120551D BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 05...
Страница 10: ...897N120551D 4 BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 06...
Страница 18: ...897N120551D 1 8 Chapter 1 Precautions BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 06...
Страница 32: ...897N120551D 3 8 Chapter 3 Name and Function of Each Part BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 12...
Страница 52: ...897N120551D 6 10 Chapter 6 Method of Use BL7000_E 50_897N120551D indb 10 2016 09 05 14 06 17...
Страница 66: ...897N120551D 8 6 Chapter 8 Troubleshooting BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 19...