44
6.2 O
2
conversion reference value setting
With the “
” indication flashing, press the CAL key
and “
” will be indicated.
The sub indication now flashes.
Set a reference value (1 to 19) %O
2
.
Press the
key and the numeric will be incremented.
Press ENT key after setting a reference value.
O
2
converted concentration
21 - Set value
=
×
NO
x
concentration
21 - O
2
concentration
Note: The maximum value to calculation is 4 .
( )
NO
CAL
ENT
∧
NO
NO
NO
6.1 Response time setting
Press CAL key while “
” display lamp is flickering,
and “
” will be displayed. At this time, the sub display
lamp flickers.
Press COMP key to select desired setting component.
Next, set the response time of the electric system.
Numerical value (1-199, O
2
: 1-49) is set from the top digit
displayed by the flicker of sub display lamp.
90% response time (electric system) ·=. 0.22
×
(1-199)
sec.
Numerical value increases by pressing
key.
Digits are selected by pressing
key.
After setting the response time, press ENT key.
( )
NO
CAL
ENT
∧
>
NO
NO
NO
(
(
Note