12
COMP1, COMP 2 input/output terminal block
<Instantaneous value output>
Instantaneous value of 0 to 1V DC or 4 to 20mA DC is outputted.
<Moving average output> (option)
Specified 1 or 4 moving average value of 0 to 1V DC or 4 to 20mA DC is outputted.
<Upper limit alarm contact output> (option)
When signal exceeds upper limit, terminals
⑤
and
⑥
turn from ON to OFF and
⑥
and
⑦
turn from OFF
to ON.
1c contact 250V AC, 2A (resistive load)
<Lower limit alarm contact output> (option)
When signal is below lower limit, terminals
⑧
and
⑨
turn from ON to OFF and
⑨
⑩
turn from OFF to
ON.
1c contact 250V AC, 2A (resistive load)
<Range identification signal output> (option)
Terminals
⑬
and
⑭
are conductive when 1st range is selected:
⑬
and
⑭
are open when 2nd range is
selected.
1a contact 250V AC, 2A (resistive load)
<Remote range input> (option)
1st range is selected when 5V DC is inputted to terminals
⑲
and
⑳
; 2nd range is selected when there is
no input to terminals
⑲
and
⑳
.
<External hold input> (Input to COMP1 terminal) (option)
Hold setting component is outputted and held with 5V DC inputted between
⑲
and
⑳
.
<Fault> (Input to COMP1 terminal)
Contact output when analyzer incurs an abnormality.
1a contact 250V AC, 2A (resistive load)