TN5A4217b 55/58
9.3
Calibration
This section describes the calibration procedures for NRF54.
Expose NRF54 to the gamma-ray sources such as
137
Cs and
60
Co and beta-ray sources suc
h as
90
Sr/
90
Y.
A dose should be measured by placing the source at a certain distance (calibration distance)
from reference point of NRF54 so that true value of the dose is traceable to the National
Standard.
(1) Determination of a reference dose value (R
0
)
-
Determine a reference dose value (R
0
) by the following method:
a. Calculate R
0
from the reference source activity, the distance between the reference
source and reference point of NRF54 (calibration distance) and irradiation time.
b. Or the dose rate value at the reference point may be simply well-known by field
calibration/characterization. In this case, reference dose value (R
0
) can be calculated
by the known dose rate value and irradiation time.
(2) Dose value (R
1
) measurement
-
Place the source at the calibration distance from reference point of NRF54.
-
Take the dose reading (R
1
) after irradiation which gives enough statistical stability.
(e.g. 1 mSv)