TN5A4217b 30/58
4.4.4 Infrared communication function
4.4.4.1 Instructions for use
(1) NRF54 should be placed so as that infrared window of NRF54 is opposite to Infrared.
(See P.8 and P.9)
(2) Setting parameter can be changed by user with Dosimeter Configuration Tool.
4.4.4.2 Specification
Communication standard: IrDA ver. 1.2 (Low Power)
Recommended infrared device: ACT-IR220LN9.6 (Made by ACTiSYS)
4.4.4.3 Icon display
Infrared icon
(1) Infrared icon turns on: Active (connection established)
(2) No icon: OFF
For infrared communication, be sure to communicate with one dosimeter at a
time. Do not communicate with more than one dosimeter at a time, such as
arranging multiple dosimeters.
Also, do not communicate near objects that reflect infrared rays, such as
mirrors.
Otherwise, data may be written to other dosimeters by mistake, a data error
may occur, or communication may become impossible.