TN5A4217b 44/58
6.3
During use (When alarm is generated)
When some alarms are generated, LCD displays are as follows:
*See Chapter 4 for operation of buzzer, vibrator and LED/Backlight during alarm generation.
No
Item
Display
Remarks
1 Accumulated dose
alarm/warning
Alarm
.
Overload
.
Warning
.
Warning alarm is
generated when
exceeding dose
warning set value.
Alarm is generated
when exceeding dose
alarm set value.
Also, display is
changed when
overload happens.
For Hp(0.07)
alarm/warning, Hp07 is
displayed instead of
Hp10.
2 Dose rate alarm/warning
Alarm
.
Overload
.
Warning
.
Warning alarm is
generated when
exceeding dose rate
warning set value.
Alarm is generated
when exceeding dose
rate alarm set value.
Also, display is
changed when
overload happens.
Alarm is canceled
when dose rate
reaches down at 80%
of warning/alarm value
or less.
For Hp(0.07)
alarm/warning, Hp07 is
displayed instead of
Hp10.
3 Operation time alarm
.
When exceeding
operating time alarm
set value, time alarm is
generated.