8. Instruction for Use
Rev. Date
07.09..2017
Page
10/13
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by one or more of
the following measures:
-- Reorient or relocate the receiving antenna.
-- Increase the separation between the equipment and receiver.
-- Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
-- Consult the dealer or an experienced radio/TV technician for help.
Should you have any inquire, pls feel free to contact me.
Guidance and manufacture’s declaration – electromagnetic emission
The eB-G-BT is intended for use in the electromagnetic environment specified below. The customer of the user of the
eB-G-BT should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The
eB-G-BT
use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class B
The
eB-G-BT
is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Guidance and manufacture’s declaration – electromagnetic immunity
The eB-G-BT is intended for use in the electromagnetic environment specified below. The customer or the user of
eB-G-BT should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with
synthetic
material,
the
relative humidity should be at
least 30%.
Power
frequency
(50Hz/60Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.