2. Preventive measure before using
(1) Wash hands with clean water and soap and use the alcohol and sponge to disinfect.
(2) Ensure that the needle is new and do NOT use the used one.
3. Measures as per the measuring result
(1) If the result is regarded not to meet the real condition of your body, please check whether the operation is
correct. And test again by the effective test strip, if the result is still same, please consult your physician in
time. Do NOT make change of taking medicine in accordance with the result only.
(2) Regularly compare with the measurement made in hospital is necessary, if you feel the symptom is not
consistent with the results. Please consult your physician for further best treatment.
4. Factors of affecting the accuracy of result, applicable scope and precautions
(
1
)
Factors of disease affecting the result:
(a) The packed cell volume is smaller than 35% or bigger than 50%.
(b) The capacity of ending blood circulation obstruction blood is reduced, such as serious dehydration,
hypotension, shock and ending blood vessel diseases.
(c) The level of triglyceride is higher than 57mmol/L.
(
2
)
Factors of wrong operation:
(a) The blood volume is not enough, the reaction is not full;
(b) The initial blood sample is not enough, measuring the test after adding the blood again;
(c) The blood sample is polluted;
(d) The test strip is pulled out and then inserted;
(e) It’s stored in the low temperature (lower than 10
℃
(50
°F
)
) or measuring with the deep frozen strip;
(f) Iodine or chlorine disinfection agent is used;
(g) Take blood from the wet finger after being disinfected;
(h) It is placed in the high temperature environment longer than 2 minutes after being unpacked.
(
3
)
Why the result of testing the ending blood is not consistent with the value of testing the venous blood glucose
in hospital? Generally, the venous blood test in hospital it take will have 15% difference if compared to the
ending blood test .
1. The blood volume is not enough which shall make the value of testing lower.
2. Pressing the bleeding area shall make a large amount of tissue fluid get into and make the thickness
of blood glucose reduced.
3.The Code card is not used or the password is not consistent with the Code on bottle.
4. The test strip is out of its service life or the bottle has been opened for more than 3 months or longer; the test
strip is damped or polluted;
5.When the alcohol is used for disinfection, the alcohol is not dry or the bleeding area is damped.
6. The iodine is used to make the value higher;
7.The ambient temperature is not proper (the best temperature should be in the range of 10
℃
-40
℃
(50 °F
~
104 °F))
。
8. Other factors: the capacity of blood corpuscle is bigger than 60%, other glucides and internal substances,