- 4 -
temperature unit
“
” or “
” blink it’s ready to measure, please see on( Figure 5.4)
●
Insert the IR Thermometer probe into the ear canal as on the following (Figure 5.1); Pull
the ear slightly back ward to straighten the ear canal. Then insert the probe into the ear as
far as it goes. Be sure the probe fully seals the ear canal. Hold this position until the
measurement is completed.
●
Press the “ SCAN “ button key to wait about 1 second then you can hear the long “Bi--”
sound that it’s finished the measurement in the meantime the measured temperature data
appears on the displayed screen. (Figure 5.5)
●
Remove the IR thermometer from the ear canal and read measured data on the displayed
screen. If you don’t use again, the power will be automatically switch off within 60 seconds.
Figure 5.1 Figure 5.2 Figure 5.3 Figure 5.4 Figure 5.5
2) Contact Type Forehead Measuring mode (Figure5.1A
)
(Suitable for FT-F21
,
FT-F22 &
FT-F21-BT,FT-F22-BT, UO8080E)
●
Turn on the “power” button key of the IR Thermometer and then all displayed segments
appear briefly as like on(Figure 5.2A),the displayed screen appears last measured
temperature data, please see on (Figure 5.3A),after few seconds automatically change to
the Forehead Measuring Mode “ ” symbol appears on the displayed screen and the
temperature unit “ ”or “
” blink then it’s ready to measure, please see on (Figure
5.4A).
●
Put the probe in the middle of forehead position and press the “SACN” button key then
keep moving the IR thermometer from the middle of forehead to the side of forehead then
you will hear long “Bi--” sound that It is finished the measurement in the meantime the
measured temperature data appears on the displayed screen, the measuring time takes
about 1 second .(see on Figure 5.5A).
●
Remove the IR thermometer from the forehead and read measured data on the displayed
screen. If you don’t use again, the power will be automatically switch off within 60
seconds.
Figure 5.1A Figure 5.2A Figure 5.3A Figure 5.4A Figure
5.5A