Section 7
Parameter Settings
68
FT742-SM (RS422 & RS485) Sensors
– User Manual
7.4.25 UC.3: Set User Calibration Table Record
Command
Parameter
UC (set & verify record)
Command
Syntax
SET Sensor Calibration
Record:
$<listenerID>,UCW<Cspeed>,<Uspeed>*<checksum><c
r><lf>
$aa,UCWxx.xx,yy.yy*hh<cr><lf>
Verify Last Record:
$<listenerID>,UC?W*<checksum><cr><lf>
$aa,UC?W*hh<cr><lf>
Sensor output:
$<talkerID>,UC=<error code>*<checksum><cr><lf>
$aa,UC=n*hh<cr><lf>
Parameters
<
Cspeed
>
xx.xx
Corrected speed
<Uspeed>
yy.yy
Uncorrected speed
<error code>
0
1
2
3
4
5
Table entry accepted
Error: Sensor speed out of order (latest row speed <previous
row speed)
Error: Sensor speed increment less than 0.5ms than previous
record
Error: Data entry not allowed (table has not been cleared first)
Error: Bad argument (data format not valid)
Error: User calibration table is full (all 64 rows have been
entered)
Examples
Example 1
Enter user calibration table record and verify
Message
Comment
$01,UCW00.90,01.11*48<cr><lf>
Set a wind speed correction
$01,UC?W*53<cr><lf>
Query if table entry was accepted
$WI,UC=0*29<cr><lf>
Sensor response
Description
Use the UCW command to set and verify individual user calibration table records. New
records can only be entered if the Calibration table is cleared first (see Section 7.4.24)
Up to 64 records can be entered sequentially into the sensor’s RAM and verified. Once
sufficient records have been loaded, these can be saved to the Flash using the user
calibration save command (see Section 7.4.26).