8
Environmental Conditions for Operation and Storage
Temperature range within 0°C to 40°C(operation), -20°C to 60°C (storage)
Relative humidity range 10% to 85%
Atmospheric pressure range within 500 to 1060hPa.
On Installation
1. Openings in the medical monitor cabinet are provided for ventilation. To prevent overheat-
ing, these openings should not be blocked or covered. If you put the medical monitor in a
bookcase or some other enclosed space, be sure to provide adequate ventilation.
2. Do not expose the medical monitor to rain or use it near water. If the medical monitor acci-
dentally gets wet, unplug it and contact an authorized dealer immediately. You can clean the
medical monitor with a damp cloth if necessary, but be sure to unplug the medical monitor
fi rst.
3. Place your medical monitor near an easily accessible AC outlet.
4. High temperature can cause problems. Max operating temperature is 40°C. Don’t use your
medical monitor in direct sunlight and keep it away from heaters, stoves, fi replaces,and
sources of heat.
5. Don’t place your medical Monitor on an unstable stand, Medical monitor may malfunction or
fall.
6. This medical monitor should not topple over when tilted at a 5° angle, in any position, during
NORMAL USE, excluding transport.
7. In the position specifi ed for transport, medical monitor shall not overbalance when tilted at a
10 degree angle.
8. When carrying this product, please use both handles (if included) on the left and right side of
the product, and carry using two people. If you want the product to be installed in another
place, please call your service center.
9. Always use only the original cables and accessories with the device.
10. Do not lay this monitor on other equipment.