3. For the use of ME devices in professional healthcare facilities.
Test specifi cation for ENCLOSURE PORT IMMUNITY to RF wireless communica-
tions equipment (according to IEC 60601-1-2:2014)
The medical monitor is intended for use in the electromagnetic environment specifi ed below.
The user of the medical monitor should make sure that it is used in such an environment.
Test fre-
quency
MHz
Band
MHz
Service
Modulation
Maximum
power
W
Distance
m
IMMUNITY
TEST LEVEL
V/m
385
380 to 390
TETRA 400
Pulse modulation
18 Hz
1.8
1.0
27
450
430 to 470
GMRS 460,
FRS 460
FM
± 5 kHz stroke
± 1 kHz sine wave
2
1.0
28
710
704 to 787
Band 13, 17
Pulse modulation
217 Hz
0.2
1.0
9
745
780
810
800 to 960
GSM 800/900
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
18 Hz
2
1.0
28
870
930
1720
1700 to
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE Band 1,3,
4, 25 UMTS
Pulse modulation
217 Hz
2
1.0
28
1845
1970
2450
2400 to
2570
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
217 Hz
2
1.0
28
5240
5100 to
5800
WLAN 802.11
a/n
Pulse modulation
217 Hz
0.2
1.0
9
5500
5785
*Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the medical monitor may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.