Real-Ear Measurements
135
If the instrument didn’t successfully level, there are several things you can try.
• Check the position of the sound field speaker. It should be 12 to 15 inches
from the client’s head.
• Make sure the microphones are securely plugged into the remote module.
• Make sure the front of the reference microphone is as close to the client’s
head as possible.
• Check the calibration of the loudspeaker and microphones. See Appendix
B.
• Check the noise level in the room. It may be necessary to change or cor-
rect the sound field environment.
See Section 2.8 for details on the leveling status and troubleshooting the level-
ing process.
6.2.5 Setting up Body Aids
The setup below is recommended for Body aids.
REFERENCE MIC
PROBE MIC
TOP VIEW
12"
SPEAKER
REFERENCE MIC
SIDE VIEW
PROBE MIC
(ON EARHOOK)
BODY AID
(FACING FORWARD)
SPEAKER
(HEAD HIGH)
12"
Figure 6.2.5—Setting up a body aid for real-ear testing
6.3 Real-Ear Targets
This section describes how to enter a real-ear audiogram, enter or measure a
real-ear to coupler difference (RECD), and create or modify a real-ear target.
6.3.1 Entering an Audiogram
1. Enter the Real-Ear Navigation screen by pressing [F2] from the Opening
screen or [EXIT] from any real-ear screen.
2. Press [F1] to enter the Audiogram Entry screen. See Figure 6.3.1.
3. Press [F1] to select the desired ear.