Real-Ear Measurements
129
12"
Speaker
(45 degrees)
Reference Mic
Probe Mic
12"
Reference Mic
Probe Mic
Top View
Front View
Speaker
(45 degrees)
Figure 6.2.1—Sound field speaker setup
6.2.2 Placing the earhook and reference microphone
1. Place the wedge-style earhook on the client’s ear.
2. Attach the reference microphone, facing forward, on the wedge of the
earhook, directly above the ear to be tested. The reference microphone
should be as close to the head as possible. See Figure 6.2.2.
3. Adjust the round Velcro pad on the wedge style earhook by turning it, so
that the inside pad fits firmly against the client’s neck.
Figure 6.2.2—Placing the reference and probe microphones