Automated Test Sequences
105
Figure 4.4—Attack & Release Test
1. Attack graph
2. Release graph
3. Measured Attack & Release times and test settings.
4. Currrent Attack & Release settings
By default, the Attack & Release test uses the settings specified by the ANSI 96
test sequence at 2000 Hz. If you want to perform attack and release measure-
ments using a different standard, use the [F4] key. The available test selections
are: IEC Loud, IEC Speech, JIS, IEC, and ANSI 96. (The ANSI 03 test sequence
uses settings identical to ANSI 96.) This changes the test settings to those speci-
fied by the selected standard.
You can further customize the test settings by pressing [MENU] to open the
local menu. Use the arrow keys to make the desired adjustments, and use
[MENU] again or [EXIT] to close the menu. See the Technical Details section
below for more information.
In addition to the settings adjustable in the local menu, you can also change the
frequency and the source type of the test. Use [F5] to toggle between Puretone
and Composite. Use the [
<, >
] keys to change the selected frequency. (You can
only change the frequency when the Source Type is set to Puretone with [F5].)
Please note that manufacturer specifications normally assume a puretone source
type.