22
Click [Next >]
The summary of the calibration process is displayed.
23
Define the calibration class:
o Precision
o Standard
24
Set the specified measured value M1 on the power source
25
Click [ Start measuring process ]
26
Enter the value displayed on the power source in the Display [A] box
27
Click [Next >]
28
Repeat steps 25 - 28 for the measured values M2 - M5
(5 measurements)
When the calibration process is complete, a corresponding notification is dis-
played and the [Print calibration report] button is available.
29
Click [Print calibration report]
Calibration system / create report
is displayed.
30
Enter the details to create the calibration report:
Under Template, select the language for the calibration report from the list
Enter the output path
Enter or search for ( [Browse] ) the file name
If the Signotec signature pad is connected and has been detected by the cal-
ibration system, a notification to this effect is displayed under Signature pad.
If not, check the connection of the Signotec signature pad to the calibration
system and then click on [Reconnect].
31
Sign the Signotec signature pad with the corresponding digital pen
32
Press [OK] on the Signotec signature pad
The signature is sent to the calibration software.
x ... Cancel signature
Refresh ... new signature
A corresponding notification is displayed after the signature has been re-
ceived successfully.
33
Click [Create PDF]
IMPORTANT!
Only after the confirmation "PDF created successfully!" is dis-
played, will the calibration report be saved.
81
EN
Содержание Calibration system 2.0
Страница 1: ...Operating instructions Calibration system 2 0 42 0426 0270 EN 012 14032022 EN Operating instructions...
Страница 2: ......
Страница 13: ...General information 13...
Страница 14: ...14...
Страница 22: ...22...
Страница 23: ...Controls connections and mechan ical components 23...
Страница 24: ...24...
Страница 28: ...28...
Страница 29: ...Installation 29...
Страница 30: ...30...
Страница 38: ...38...
Страница 39: ...Calibrating welding systems 39...
Страница 40: ...40...
Страница 47: ...2 2 Connect the Signotec signature pad to the calibration system 47 EN...
Страница 90: ...90...
Страница 91: ...Troubleshooting 91...
Страница 92: ...92...
Страница 95: ...Technical data 95...
Страница 96: ...96...
Страница 99: ...99 EN...