14
Decontamination
Following contamination of the device by hazardous substances, the operator must ensure that proper decontamination
is performed.
Do not use decontamination products that could cause a hazard due to the reaction with the components of the device
or the potential materials inside. If in doubt about the cleaning product, please contact Froilabo Service
.
In case of contamination of the inner chamber with hazardous biological or chemical materials, we recommend
decontamination of the interior with commercial type products.
Alternativement, vous pouvez utiliser les désinfectants suivants pour :
Inner tank (stainless steel):
•
commercial surface disinfectants without acid or halides (without drops)
•
Alcohol solutions
Door seal (silicone) :
•
Alcohol solutions
Inner wickets:
•
commercial surface disinfectants without acid or halides (without drops)
•
Alcohol solutions max. 10%
After cleaning, rinse using a sterile cloth.
Before restarting, the device must be thoroughly dried and ventilated as explosive gases may be formed during
disinfection.
During each decontamination, Pendant chaque décontamination, ensure the protection of people
adapted to the risks. DO NOT USE acid or chlorine base cleaners (corrosion risks) !
In the event of any deterioration and corrosion as a result of the use of non-recommended cleaners, FROILABO declines
all responsibility.
7.3 Long time stop
When the device is not used for a long time, it is best to do as described:
•
Stop the freezer.
•
Unplug the main.
•
Defrost, clean and dry the inner tank.
•
Leave the door open to prevent bad smells.
•
Remove valve foam and let it dry.