9
4.9 Starting up
Respect the order of the following instructions:
1. Equip the levels of drawers / shelves. Do not forget to remove any foam padding at the bottom of the shelves.
2. Connect the device to a single-phase earthed network protected by a 30 mA differential circuit breaker.
3. Press the ON / OFF button (activates 24V). A buzzer sounds, clear the alarm by pressing the button.
4. Make sure that the compressor and the fan start.
5. The display shows the first 2 pictograms (sector and set temperature not reached).
6. Set the set point (-45 ° C or -80 ° C default, depending on the model).
7. Attendre que l’appareil atteigne la température de consigne (3 à 4 heures selon le modèle). Le pictogramme de
température de consigne non atteinte disparait lorsque la température atteint -60°C (pour le modèle -80°C) ou
-30°C (pour le modèle -45°C).
8. Wait for the appliance to reach the set temperature (3 to 4 hours depending on the model). The “set temperature
not reached” pictogram disappears when the temperature reaches -60 ° C (for the -80 ° C model) or -30 ° C (for
the -45 ° C model).
9. Once the set temperature is reached, you can
charge the device
.
Nota 1:
Charging the device with "hot" products may cause the device to warm up. The high temperature
alarm could then activate.
When the device is installed, the injection must only be activated once the target temperature has
been reached, otherwise the cylinders will be unloaded during the first temperature decrease.