5.3 Setting CO

2

 / LN

injection 

To be used, the 

CO

2

 / LN

2

 

emergency injection system must be activated and the injection threshold must be set.

  

 

Configuration de l’injection CO

2

 / LN

2

1.  Press 

SEL 

to access parameters. 

2.  Press 

 to display 

CH 5

, validate by pressing 

SEL

3.  Press 

 to display 

A2tp

, validate by pressing 

SEL

4.  Enter the value 1 using the 

 key. Validate using 

SEL.

 

CO

2

 / LN

threshold is activated. 

 

5.  Go back to main menu by pressing 
6.  Press 

SEL 

1 time.  

7.  Press 

 2 times to display 

AL2

, validate by pressing 

SEL.

 

8.  Modify the triggering CO

2

 / LN

2

 temperature by using 

 and 

 keys.  

9.  Validate by pressing 

SEL

10.  Go back toi main menu by pressing 

 

 

When the device is  installed, the injection must only be  activated once the target temperature has 
been reached, otherwise the cylinders will be unloaded during the first temperature decrease. 

 

6. Indicators et alarms LED 

The main display allows a clear and visible display of indicators and warning lights. They have a simplified color code 
to help you easily understand the type of defect: 
 

 

Green: Normal running 

 

Orange: minor defect (make the necessary arrangements as soon as possible)  

 

Red: major defect (make the necessary arrangements immediatly) 

 

Maintien des fonctions alarmes : 

If  there  is  no  main  power,  alarms  (buzzer,  alarms  report  and  CO

2

/N

2

  [optionnal])  stay  activated  thanks  to  the  24V 

batteries supplied as standard with your freezer. Temperature display is also maintained by the 24V batteries.  
 

 
Warning: compressors are no longer powered, so the freezer doesn’t produce cold anymore. 

 

 

When a 

major alarm 

is triggered, the 

« OK » 

indicator on the screen disappears until the problem is resolved.  

 

6.1 Main power connected 

This icon appears when the 230 VAC power supply is connected. It disappears when the 230 VAC power 
supply is disconnected. 

If the alarm is cut off via the front panel button, it will restart after 

30 minutes

.  

 

6.2 Temperature alarm (high or low) 

This  light  indicates  that  the  set  temperature  inside  the  freezer  is  not  reached.  Check  the  setting  on  the 
regulator or contact after sales service. 

If the alarm is cut off via the front panel button, it will restart after 

4 hours

.  

Содержание TRUST Series

Страница 1: ...User manual Ref UM_TRUST EN_rev1 0 ULT FREEZER TRUST 45 C 86 C...

Страница 2: ...rst use This document has been prepared with the utmost care possible However Froilabo declines all responsibility in the event of errors or omissions The same applies to any damage arising from the u...

Страница 3: ...high and low temperature alarms 10 5 3 Setting CO2 LN2 injection 11 6 Indicators et alarms LED 11 6 1 Main power connected 11 6 2 Temperature alarm high or low 11 6 3 Overpressure CP1 12 6 4 Battery c...

Страница 4: ...or recycling Companies installing servicing maintaining repairing or commissioning equipment containing refrigerants must have a certificate referred to in Article R543 76 of the French Environment C...

Страница 5: ...and operating instructions provided in that manual Froilabo SAS guarantees that the equipment will operate optimally in accordance with the conditions of installation and use set out in this manual T...

Страница 6: ...r to the corresponding safety data sheets 4 First contact with your freezer 4 1 Delivery and unpacking Use a pallet truck to move the freezer on its pallet It is imperative to keep the freezer unpacke...

Страница 7: ...moisture content in the air should ideally not exceed 50 RH The use of air conditioning can significantly extend the life of compressors If you are using a liquid CO2 or LN2 liquid backup device refer...

Страница 8: ...he equipment 4 5 Construction and isolation The one piece electro galvanized steel exterior body is protected by an epoxy paint The inner tub is made of stainless steel The thermal insulation is ensur...

Страница 9: ...ice and the calibrated foam are working properly Reposition properly after cleaning After stopping the freezer completely remove the valve foam to dry Before restarting the freezer replace the dry foa...

Страница 10: ...Maximum opening angle of the door 110 Internal dimensions without rack Height mm 716 716 1076 1076 1436 1436 Width mm 630 630 630 630 630 630 Depth mm 752 752 752 752 752 752 Power consumption in regu...

Страница 11: ...nd or drawers simultaneously The maximum permissible load for each sliding shelf or drawer must never be exceeded 75 kg These devices are not explosion proof The internal surfaces of the freezer and i...

Страница 12: ...la temp rature de consigne 3 4 heures selon le mod le Le pictogramme de temp rature de consigne non atteinte disparait lorsque la temp rature atteint 60 C pour le mod le 80 C ou 30 C pour le mod le 4...

Страница 13: ...C for 80 C model or 30 C for 45 C model Setting the high temperature alarm A1 H 1 Press SEL to access parameters 2 Hold SEL apparition of A1 L parameter 3 Display A1 H parameter by pressing 1 time and...

Страница 14: ...ified color code to help you easily understand the type of defect Green Normal running Orange minor defect make the necessary arrangements as soon as possible Red major defect make the necessary arran...

Страница 15: ...ossible This alarm does not trigger any sound signal from the freezer 6 5 Open door alarm This indicator appears when the door remains open for more than 2 minutes It disappears when closing the door...

Страница 16: ...d bleach even very diluted Never rub stainless steel with metal sponges or other abrasives Beware of the risk of burns Freezers are equipped with removable racks for easy maintenance Rack disassemble...

Страница 17: ...without drops Alcohol solutions Door seal silicone Alcohol solutions Inner wickets commercial surface disinfectants without acid or halides without drops Alcohol solutions max 10 After cleaning rinse...

Страница 18: ...he main 24V switch on 0 Switch the 24V switch to I Defective power supply Check the electrical installation Defective power cord Replace the power cord Regulator does start but the freezer doesn t pro...

Страница 19: ...and low points pits drainage channels basements where it could accumulate and render the atmosphere unbreathable Use a CO2 concentration detector or O2 concentration detector to check that the CO2 con...

Страница 20: ...n 16 oxygen is hazardous nitrogen gas displaces oxygen in the air Nitrogen gas builds up in low areas Risk of frostbite Risk of corrosion of steels in the presence of humidity In closed vessels the pr...

Страница 21: ...ratures Treat valves with care operate gently do not dismantle or lubricate valves All equipment used must be in good condition and designed for use at the intended temperature and pressure Mobile equ...

Страница 22: ...Waste Electrical and Electronic Equipment WEEE According to this directive it is now forbidden to discard industrial devices of which this product is a part delivered after 13 08 2005 with municipal...

Страница 23: ...________________________________ Kind of devices ________________________________________________________________________ Temperature __________________________________________________________________...

Отзывы: