Technical characteristics:
Characteristics
TRUST Freezers
340L
515L
690L
-86°C
-45°C
-86°C
-45°C
-86°C
-45°C
General
Raw volume (liters)
340
340
515
515
690
690
Number of compartments
2
2
3
3
4
4
Freezer weight (kg)
223
178
267
222
330
285
Power consumption* at 230 V at 50 Hz (Watts)
900
1000
1150
1000
1150
1000
Power consumption* at 110 V at 60 Hz (Watts)
1000
1100
1400
1100
1400
1100
Minimum cooling capacity of the rooom containing
the continuous freezer (Watts)
850
950
1200
950
1200
950
Sound pressure measured in dB(A) (in regulation
mode)
56
External dimensions
Height (mm)
1280
1280
1640
1640
2000
2000
Width (mm)
875
875
875
875
875
875
Depth (mm)
970
970
970
970
970
970
Required clerance at the back of the device (mm)
200
200
200
200
200
200
Maximal depth with the door open (mm)
1750
Maximal width with the door open (mm)
1150
Maximum opening angle of the door
110°
Internal dimensions without rack
Height (mm)
716
716
1076
1076
1436
1436
Width (mm)
630
630
630
630
630
630
Depth (mm)
752
752
752
752
752
752
* Power consumption in regulation mode, all components activated and stabilized, at 23 ° C (ambient temperature) with the freezer being empty.
On an equivalent product, hydrocarbon-based (HC) gases can reduce both power consumption by up to 25% and cooling requirements.
Refrigeration system:
Characteristics
TRUST Freezers
340L
515L
690L
Standard
gazes
Green
gazes
Standard
gazes
Green
gazes
Standard
gazes
Green
gazes
Double stage -86°C models
Hermetic compressor power stage 1 (Watt)
580
380
780
580
780
580
Hermetic compressor power stage 2 (Watt)
380
380
580
580
580
580
Stage 1 R417a refrigerant– load(g)
600
/
600
/
600
/
Stage 2 R508b refrigerant– load (g)
300
/
300
/
300
/
Stage 1 R290 refrigerant– load (g)
/
140
/
200
/
200
Stage 2 R170 refrigerant– load (g)
/
190
/
80
/
90
Simple stage -45°C models
Hermetic compressor power
380
/
580
/
780
/
ISCEON 89 refrigerant – load (g)
600
/
600
/
600
/
Capillary relaxation
YES
Air condenser as stantard
YES