4.5 Ambient conditions
WARNING!
Mains voltage
– The device may only be operated indoors.
– The surrounding air must not contain any electrically
conductive particles.
– Maximum relative humidity 80% for temperatures up
to 31 °C, linearly decreasing down to 50% relative
humidity at 40 °C.
n
The room temperature should be between 5 and 40 °C.
n
Altitudes up to 2000 m
n
Degree of pollution 2 according to IEC 60664-1:2007.
4.6 Electrical connection
DANGER!
Provide short-circuit protection!
Risk of damage due to short-circuits.
– Make sure that the socket is connected to a mains line
protected with a residual current circuit breaker.
DANGER!
Mains voltage!
Changes to the connection line may only be made by a
qualified person.
CAUTION!
Ignoring the values on the type plate may result in damage
to the electrical and mechanical components.