Do not switch off the device during this work. When the
device is switched off, the sample material can stick or
glue to the sieves.
Transfer (rinse) the sieve residue into weighed bottles and dry it.
Now the lower sieves can be sieved further. Continue dismantling as
described above.
NOTICE!
The sieve stack must remain filled with liquid during the
entire sieving process.
The sieve foils must not run dry!
That can cause irreparable damage to the sieves.
Torn sieve foils can be soldered with soft solder, point by
point from underneath.
6.5 Standby
If the device is not operated for approx. one hour, it switches to energy-
saving mode (STANDBY). Only the STANDBY lamp lights up.
The device is switched to normal mode by pressing the START key. The
POWER lamp lights up, the STANDBY lamp goes out.
6.6 Setting the sieving time
The sieving time can be set in increments of minutes using the +/- keys.
Values between 1 and 99 minutes are possible; after a sieving time of 1
hour, let the device cool down.
Continuous operation is set by holding down the "-" key until "P" or "00"
appears on the display.
Using the device
- 31 -