2.7 Entering the temperature limit
1.
Select the maximum temperature input.
If the max. temperature is reached, the device stops
the grinding process and goes into cooling mode!
Depending on the setting of the operating mode, see
‘Selecting the operating mode’ on page 50, the
device stops the grinding process and opens the
grinding chamber cover or reduces the speed gradu‐
ally until a minimum speed is reached in order to ach‐
ieve an optimal cooling effect.
2.
Enter the required max. temperature limit.
DANGER!
Do not exceed a maximum temperature of 125°C. Other-
wise, the EASY GTM system and the machine will be dam-
aged.
3.
Use the [<] button to confirm your entry and return to the EASY
GTM sub-menu.
4.
Select the minimum temperature input.
When the selected min. temperature (e.g. 60 °C) is
reached again during the cooling phase, the mill
starts the grinding process again.
5.
Enter the required min. temperature limit.
6.
Press the[<] button to confirm your entry and return to the EASY
GTM sub-menu.