9.1 Replacement of the device fuse
DANGER!
Danger to life due to power surges!
Exposed power contacts
– You could touch the fuses or fuse holder with live
contacts when changing the fuses. Electric shocks
can cause burns and cardiac arrhythmia or respira-
tory and cardiac arrests.
Remove the mains cable before changing the fuses.
Always replace both fuses:
Fuse type: 2 x T10 A 250 V
1.
Press the two tabs of the fuse holder together and pull the fuse
holder out.
2.
Remove the fuses from the fuse holder and insert the new fuses.
3.
Insert the fuse holder with the inserted fuses into the opening.
9.2 Filzring Mahlraumkassette
Grease the felt ring (F) in the grinding chamber cartridge every week with
a few drops of machine oil.
If the felt ring is worn, remove the grinding set. Remove the metal ring
(M) on the rear of the grinding chamber cartridge by loosening the 3
screws and replacing the felt ring (F). Put everything back together in the
reverse order. (see Chapter
Chapter 6.3 ‘Installing the grinding ele‐
Maintenance
- 31 -