P/N 450368-03 Rev. E
3
S
ECTION
2:
S
PECIFICATIONS
A
ND
I
NSTALLATION
R
EQUIREMENTS
Prior to operating the Fresenius Medical Care GranuFlo Dissolution Unit II: 99 Gallons (375 Liters), adequate
power and water connections must be made. The GranuFlo Dissolution Unit II is a computer-controlled
electro-mechanical device. The GranuFlo Dissolution Unit II must be installed in an appropriate environment
(indoors) free from extremes of temperature and humidity, free of the presence of flammable anesthetics,
reasonably protected from dust and dirt contamination, liquid splashes, bright sunlight or high intensity
lighting.
Failure to install the GranuFlo Dissolution Unit II according to these guidelines may result in improper
operation or early failure of the device or its components. Direct sunlight may cause discoloration of the
polyethylene tank, lid, or electronic housing.
The GranuFlo Dissolution Unit II needs to be set up using the following parameters:
1.
A 15 amp
GFI
duplex outlet is required. The GranuFlo Dissolution Unit II requires 120 VAC, 60 Hz,
single phase power. The GranuFlo Dissolution Unit II uses approximately 9 amperes total. Electrical
service should meet all applicable national, state and local code requirements.
2.
The electrical power cord, input water hose, and output drain hose shall be properly routed away
from high traffic areas and installed in such a way that does not cause any electrical or safety hazards.
3.
The treated water source must meet
ANSI/AAMI
or
ISO
requirements for water for hemodialysis,
currently
ANSI/AMMI RD62
, or
ISO 13959
. Plumbing service must meet all national, state and local
plumbing codes. The GranuFlo Dissolution Unit II uses approximately 99 gallons (375 liters) of water
per batch of product.
WARNING!
DO NOT PERMANENTLY INSTALL
T
RANSFER
H
OSE TO A STORAGE TANK
/
CONTAINER
.
4.
Supply Water Treatment System should be capable of a minimum flow rate of ½ gallon per minute.
The maximum input water pressure is 60psi.
5.
The water temperature should be 20°-30°C (68°-86°F) for dissolution mixing conditions.
6.
Height from the floor to the top of the GranuFlo Dissolution Unit II is approximately 56 inches.
7.
Platform for the GranuFlo Dissolution Unit II is 50 inches long by 28 inches wide.
8.
Transfer Filter Housing is equipped with a ¾” hose barb fitting.
9.
There is a ¾” hose barb inlet fitting, and another for the drain outlet.
10.
The GranuFlo Dissolution Unit II must be level.
11.
The GranuFlo Dissolution Unit II weight is 200 lbs. empty and 1025 lbs. filled.
CAUTION!
T
HE FLOOR MUST BE CAPABLE OF WITHSTANDING THE MAXIMUM WEIGHT OF
1025
POUNDS
.
A floor drain emptying into a sewer line must be in the immediate area of the GranuFlo Dissolution Unit II.
NOTE:
T
HE
G
RANU
F
LO
D
ISSOLUTION
U
NIT
II
T
ANK IS MADE OF NON
-
FERROUS MATERIAL
.
Содержание GranuFlo 450368-03
Страница 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 39: ...P N 450368 03 Rev E 33 PAGE INTENTIONALLY LEFT BLANK ...
Страница 41: ...P N 450368 03 Rev E 35 PAGE INTENTIONALLY LEFT BLANK ...
Страница 48: ...42 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 49: ...P N 450368 03 Rev E 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Страница 50: ...44 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 51: ...P N 450368 03 Rev E 45 ...
Страница 52: ...46 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 53: ...P N 450368 03 Rev E 47 ...
Страница 54: ...48 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 55: ...P N 450368 03 Rev E 49 ...
Страница 56: ...50 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 57: ...P N 450368 03 Rev E 51 ...
Страница 58: ...52 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 59: ...P N 450368 03 Rev E 53 Appendi x C Dry Acid Mixing Instructions ...
Страница 60: ...54 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 61: ...P N 450368 03 Rev E 55 ...
Страница 62: ...56 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 63: ...P N 450368 03 Rev E 57 ...
Страница 65: ...P N 450368 03 Rev E 59 PAGE INTENTIONALLY LEFT BLANK ...
Страница 67: ...P N 450368 03 Rev E 61 PAGE INTENTIONALLY LEFT BLANK ...