7 . 1 S P E C I F I C G R A V I T Y T E S T Continued
16
P/N 450368-03 Rev. E
14.
When the FINAL WATER LEVEL is reached, the unit will
automatically step to HOMOGENIZE Operation. Remove the Large
Access Lid and ensure the solution level has reached the Final Fill
Sensor. Place Large Access Lid onto the GranuFlo Dissolution Unit II .
Then, CLOSE the Water Supply Valve to the GranuFlo Dissolution
Unit II .
15.
During HOMOGENIZE Operation, the Mixer will mix the solution for
ten (10) minutes. When the HOMOGENIZE Operation is complete,
the Transfer Indicator
LED
will flash. Remove Large Access Lid and
look into the tank to make sure the granules have dissolved and the
solution is colorless.
16.
Once you have verified the granules are dissolved and the solution is
colorless, the solution is ready for
Specific Gravity
Testing. If the granules
have not dissolved and/or the solution is not colorless, call Technical
Service (1-800-227-2572) for further instructions.
7.1.
T
RANSFER
H
OSE
F
LUSH AND
S
PECIFIC
G
RAVITY
T
EST
:
Once the Transfer Indicator Light flashes, the concentrate is ready to be tested. This test is done using a
hydrometer which measures the density of the acid concentrate. The resultant value is called the "
specific
gravity
". The
specific gravity
is measured to verify that the concentrate has been properly mixed.
W
ARNING
!
A
LWAYS CHECK THE
S
PECIFIC
G
RAVITY OF EACH NEW BATCH OF
A
CID
C
ONCENTRATE BEFORE USING IN
TREATMENT
.
S
UPPLYING AN IMPROPERLY MIXED
A
CID
C
ONCENTRATE TO THE
H
EMODIALYSIS
M
ACHINE MAY
RESULT IN PATIENT INJURY OR DEATH
.
To determine a
Specific Gravity
Reading:
1.
Prepare the Hydrometer and the Hydrometer Cylinder:
•
Check the
Hydrometer
for cracks. Do not use a cracked
Hydrometer
.
•
Always handle the Hydrometer with care. Always hold it vertically by the top, as finger marks lower
down can affect the accuracy of the instrument.
•
The Hydrometer should never be held by the stem horizontally.
•
Rinse the hydrometer and the hydrometer cylinder separately with purified water before checking
solution for specific gravity. The purified water source must meet
ANSI/AAMI
or
ISO
standards for
dialysis currently
ANSI/AAMI RD62
, or
ISO 13959
.
•
Once
Hydrometer
is clean, place the
hydrometer
onto a clean area.
NOTE:
D
O NOT PLACE
A
H
YDROMETER
INTO THE
G
RANU
F
LO
D
ISSOLUTION
U
NIT
II
AT ANY TIME
.
W
ARNING
:
S
PECIFIC
G
RAVITY
T
EST MUST BE PERFORMED AFTER EVERY BATCH OF CONCENTRATE THAT IS MADE
.
I
N
ADDITION
,
CONCENTRATE
SAMPLES SHOULD BE TAKEN FROM THE
T
RANSFER
H
OSE ONLY
.
Содержание GranuFlo 450368-03
Страница 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 39: ...P N 450368 03 Rev E 33 PAGE INTENTIONALLY LEFT BLANK ...
Страница 41: ...P N 450368 03 Rev E 35 PAGE INTENTIONALLY LEFT BLANK ...
Страница 48: ...42 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 49: ...P N 450368 03 Rev E 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Страница 50: ...44 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 51: ...P N 450368 03 Rev E 45 ...
Страница 52: ...46 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 53: ...P N 450368 03 Rev E 47 ...
Страница 54: ...48 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 55: ...P N 450368 03 Rev E 49 ...
Страница 56: ...50 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 57: ...P N 450368 03 Rev E 51 ...
Страница 58: ...52 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 59: ...P N 450368 03 Rev E 53 Appendi x C Dry Acid Mixing Instructions ...
Страница 60: ...54 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 61: ...P N 450368 03 Rev E 55 ...
Страница 62: ...56 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 63: ...P N 450368 03 Rev E 57 ...
Страница 65: ...P N 450368 03 Rev E 59 PAGE INTENTIONALLY LEFT BLANK ...
Страница 67: ...P N 450368 03 Rev E 61 PAGE INTENTIONALLY LEFT BLANK ...