M A I N T E N A N C E Continued
26
P/N 450368-03 Rev. E
8.4.
U
NIT
F
ILTER AND
B
ASE
S
TAND
P
IPE
F
ILTER
M
AINTENANCE
The 1-micron filter (FMCNA, P/N
G84-202-12)
should be changed under the following conditions:
1. The filter should be changed after mixing 6 batches.
2. When the GranuFlo Dissolution Unit II requires disinfection.
8.4.1
.
F
ILTER
R
EMOVAL
A
ND
R
EPLACEMENT
Ensure the Mixing Tank is empty and that power to the GranuFlo Dissolution Unit II is OFF and the Main
Transfer Ball Valve is CLOSED. Follow steps A, B, C, and D for removal and replacement of the filter (Figure
19). If the residual fluid in the Filter housing is concentrate, then place the solution in the
Residual
Solution Bucket.
Figure 19
8.4.2.
B
ASE
S
TAND
P
IPE
F
ILTER
R
EMOVAL
Using Base Filter Rod, P/N 260025 (Figure 20A), reach into the tank and connect the small end of the
shaft into the Screw Nut of Base Filter (Figure 20B). Turn the Screw Nut Counter Clockwise until the Base
Stand Pipe Filter is no longer attached to the tank. Use the Base Filter Rod to help lift the Base Stand Pipe
Filter out of Tank. Thoroughly clean off the top of the Pie Plate Filter holes by running purified water over
the filter holes until all debris is removed. Once the Pie Plate Filter is clean reinstall the Base Stand Pipe
Filter.
F
IGURE
20
NOTE:
T
HE FILTER USED MUST BE COMPATIBLE WITH
F
RESENIUS
M
EDICAL
C
ARE
D
RY
A
CID
P
RODUCT AND RATED AT
1
MICRON
.
F
RESENIUS
M
EDICAL
C
ARE
,
PN
G84-202-12,
IS A
P
OLYPROPYLENE FIBER WOUND ON A POLYPROPYLENE MESH CORE MEETS THESE
REQUIREMENTS
.
C
ELLULOSE FILTERS ARE NOT COMPATIBLE WITH THE
F
RESENIUS
M
EDICAL
C
ARE
D
RY
A
CID
P
RODUCT
.
NOTE:
I
T IS RECOMMENDED TO REPLACE THE FILTER AFTER MIXING
6
BATCHES OR IF THE TANK REQUIRES DISINFECTION
.
A. Close Main Transfer Ball Valve
and Remove Filter Housing
C. Insert Filter
(P/N G84-202-12)
D. Install Filter Housing onto
the filter holder.
B. Remove Filte
r
A. P/N 260025, Base Filter Rod
B. Base Stand Pipe Filter Removal
Figure 20
Содержание GranuFlo 450368-03
Страница 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Страница 39: ...P N 450368 03 Rev E 33 PAGE INTENTIONALLY LEFT BLANK ...
Страница 41: ...P N 450368 03 Rev E 35 PAGE INTENTIONALLY LEFT BLANK ...
Страница 48: ...42 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 49: ...P N 450368 03 Rev E 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Страница 50: ...44 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 51: ...P N 450368 03 Rev E 45 ...
Страница 52: ...46 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 53: ...P N 450368 03 Rev E 47 ...
Страница 54: ...48 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 55: ...P N 450368 03 Rev E 49 ...
Страница 56: ...50 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 57: ...P N 450368 03 Rev E 51 ...
Страница 58: ...52 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 59: ...P N 450368 03 Rev E 53 Appendi x C Dry Acid Mixing Instructions ...
Страница 60: ...54 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 61: ...P N 450368 03 Rev E 55 ...
Страница 62: ...56 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Страница 63: ...P N 450368 03 Rev E 57 ...
Страница 65: ...P N 450368 03 Rev E 59 PAGE INTENTIONALLY LEFT BLANK ...
Страница 67: ...P N 450368 03 Rev E 61 PAGE INTENTIONALLY LEFT BLANK ...