Hydraulic Component Descriptions (cont.)
111 – bi
b
ag Air Separation Chamber
The bi
b
ag air separation chamber separates air from the sodium bicarbonate concentrate
upon leaving the bi
b
ag. It also is used to separate air from the sodium bicarbonate
concentrate supplied by external sources (pre-mixed concentrates).
112 – bi
b
ag Air Separation Chamber Air Sensor
The bi
b
ag air separation chamber air sensor detects air in the air separation chamber.
113 – bi
b
ag Conductivity Cell
The bi
b
ag conductivity cell is used to measure the conductivity of the sodium bicarbonate
concentrate leaving the bi
b
ag and the conductivity of the pre-mixed concentrates.
114 – bi
b
ag Temperature Thermistor
The bi
b
ag temperature thermistor is used to measure the temperature of the bicarbonate
concentrate leaving the bi
b
ag and the pre-mixed concentrate.
115 – bi
b
ag Present Switch
The bi
b
ag present switch is built into the bi
b
ag connector. The switch is positioned so
that when a bi
b
ag is attached to the bi
b
ag connector the switch is pressed indicating the
presence of a bi
b
ag bag.
116 – Bicarbonate Temperature Thermister
Used with conductivity cell 117 to measure conductivity.
117 – Bicarbonate Conductivity Cell
Measures conductivity of the bicarbonate concentrate from the bi
b
ag after it is mixed
with R.O. water.
118 – bi
b
ag Filter
Removes any particles that may enter through the bi
b
ag.