Page 18
bi
b
ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 14 (15) –
bi
b
ag
(cont.)
Fill
System is running the bi
b
ag fill process (Initial state).
1 =Yes / 0 = No
Bic Pump
System requests the bicarbonate pump to be on for a bi
b
ag process.
1 =Yes / 0 = No
Bypass Ctl
System requests bypass of the dialyzer for a bi
b
ag process.
1 =Yes / 0 = No
V43 Ctl
System is suppressing the “Valve 43 Failure” error for a bi
b
ag process.
1 =Yes / 0 = No
TMP Ctl
System requests TMP control, which isolates the dialyzer and freezes dialysate pressure
for a bi
b
ag process.
1 =Yes / 0 = No
Vent En
System indicates that it is enabled to do the bi
b
ag vent process.
1 =Yes / 0 = No
Air
Air is detected by the bi
b
ag air separation chamber air sensor.
1 =Yes / 0 = No
Empty
System is running the bi
b
ag empty process.
1 =Yes / 0 = No
Emptied
System indicates that the bag is emptied.
1 =Yes / 0 = No