Page 26
bi
b
ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 15 (16) – Bic Mon & Act. Reg
(cont.)
V104 Er O
Indicates Valve 104 or Valve 108 is Stuck Open.
1 =Yes / 0 = No
V104 Er C
Indicates Valve 104 or Valve 108 is Stuck Closed.
1 =Yes / 0 = No
Acid PresOff
The compensation value for Acid Port Pressure Transducer 106 based on the Regulator
Pressure service mode calibration.
Bic PresOff
The compensation value for bi
b
ag Pressure Transducer 110 (bi
b
ag machine) based on the
Regulator Pressure service mode calibration.