Page 24
bi
b
ag V2.0 Technician’s Manual
P/N 490188 Rev. D
Debug Screens (cont.)
Debug Screen 15 (16) – Bic Mon & Act. Reg
(cont.)
RO Cond
The RO water conductivity value measured at the end of rinse used to compensate “Bic
Cell Cond”.
TCB
Theoretical Conductivity of Bicarbonate based on the machines Base Na+ and
Bicarbonate settings.
Bic Lo Th
The lower threshold for compensated bicarbonate conductivity (Bic Mon Cond), 5%
below TCB.
Bic Mon Cond
The compensated bicarbonate conductivity based on Bic Cell Cond, Bic Mon temp, Bic
Slope, Bic Offset, and RO Cond.
Bic Hi Th
The upper threshold for compensated bicarbonate conductivity (Bic Mon Cond), 5%
above TCB.
Bic Slope
The slope value of the bicarbonate conductivity calculated from the bicarbonate
conductivity cell calibration.
Bic Offset
The offset value of the bicarbonate conductivity calculated from the bicarbonate
conductivity cell calibration.
Acid Press
The compensated pressure seen at the Acid Port Pressure Transducer 106.
Bic Press
The compensated pressure seen by the bi
b
ag Pressure Transducer 110 (bi
b
ag machine) or
the bicarbonate Pressure Transducer 107 (non-bi
b
ag machine).
Inlet W temp
– (Not used)