PROGRAMMING
The following program functions are always available to view and change:
P1 - Software Program Revision Number - Read Only
P2 - Product Manufacturing Date - Read Only
P3 - Set Current Date - Read/Write
P4 - Set Current Time - Read/Write
P5 - Retrieve Fault Codes - Read Only
Access Program Features
Note:
When the program (P) code is flashing in the
PUMP INTAKE
display,
press the
INC
or
DEC
button to scroll through the P-codes or press the
SILENCE
button to exit the programming mode.
1. Press the
SILENCE
button and hold it until the
RPM
display shows four dashes
– – – –
and the message display shows
ENTER---
. Release the button.
Result:
P 1
flashes in the
PUMP INTAKE
display. The message display shows the
program revision number
PROG REV V100.03
.
2. Press the
INC
button.
Result:
P 2
flashes in the
PUMP INTAKE
display. The message display shows the
manufacturing date
MFG DATE 09AUG'07
(ddmmm'yy).
3. Press the
INC
button.
Result:
P 3
flashes in the
PUMP INTAKE
display. The message display shows the
current date
SET DATE 17AUG'07
.
4. To Change the Date: (If not go to step 5.)
a. Press the
MENU
button.
Result:
P 3
stops flashing. The message display shows the current date
with the
year flashing.
b. Press the
INC
or
DEC
button to change the year.
c. Press the
MENU
button.
Result: The month flashes.
d. Press the
INC
or
DEC
button to change the month.
e. Press the
MENU
button.
Result: The day flashes.
f. Press the
INC
or
DEC
button to change the day.
g. Press and hold the
SILENCE
button to store the new date.
Result:
P 4
flashes in the
PUMP INTAKE
display. The message display shows
SET TIME 10:30AM
. Go to step 6.