7.0 IEEE 488.2 commands
Commands
•
*CAL:
Conducts a calibration (initialization run) if the device is not calibrated. During initialization
run, the logical device cannot be changed!
•
*CAL?:
Query of calibration status of the selected logical device.
Answers:
0: calibration successful
1: calibration failed
2: calibration in course
•
*CLS:
Clear Status sets the status registers ESR, STB, OPER and QUEST at zero.
•
*ESE <0...255>:
Event Status Enable sets the event status enable register with the selected value.
•
*ESE?:
Query of the event status enable register.
•
*ESR?:
Query and deletion of the event status register.
•
*IDN?:
Query of the identification text (Frankonia, FC02, 0, V1.0).
•
*OPC:
Operation Complete sets the OPC bit (bit 0) in the event status register, after completed
execution of all previous commands.
•
*OPC?:
Query Operation Complete reports an additional 1, after completed execution of all previous
commands.
•
*RST:
Sets the equipment to initial state.
Limit:high = maximum, Limit:low = minimum, Offset = 0, Direction = up, Step = 0.1m (a
clamp) / = 10° (turntable), velocity = 10
•
*SRE <0...255>:
Service Request Enable sets the service request enable register with the selected
value.
•
*SRE?:
Query of the service request enable register.
•
*STB?:
Query of the status register (serial poll)
•
*TST?:
Equipment performs self test and reports 0 (no failure) or 1 (failure).
•
*WAI
: Wait to Continue allows the execution of a new command only after completed execution of all
previous commands.
When the controller is switched on, all status and enable registers are set to 0 and all parameters are set to
initial state (Exception: Bit 7 of the event status register is set to „Power on“.)
MA_FC02_(Rev2_1) - 9/06 / Version 1 / Rev:2.1